Vacuum Pump - Diffusion Pump – VHS-250 / VHS-10 / VHS-400 / HS-16
|
|
|
|
|
VHS-250
|
VHS-10
|
VHS-400
|
HS-16
|
Pump Model
|
VHS-250
|
VHS-10
|
VHS-400
|
HS-16
|
Previous Model Number
|
–
|
0185
|
–
|
0164
|
Speed, l/s (operating range)
|
|
|
|
|
Air
|
2,125
|
3,650
|
4,500
|
6,000
|
Helium & Hydrogen
|
2,660
|
4,560
|
5,625
|
7,500
|
Maximum forepressure, Torr (mbar)
|
|
|
|
|
No load
|
0.65 (.85)
|
0.65 (.85)
|
0.65 (.85)
|
0.65 (.85)
|
Full load
|
0.55 (0.72)
|
0.55 (0.72)
|
0.55 (0.72)
|
0.55 (.72)
|
Maximum throughput,
|
|
|
|
at 8100 W
|
T-l/s (mbar-l/s) In operating range
|
2.6 (3.5)
|
6.3 (8.4)
|
6.3 (8.4)
|
9.5 (11.7)
|
@ 1 x 10-2 Torr (1.3 x 10-2 mbar)
|
3.5 (4.5)
|
7.5 (10.0)
|
7.5 (10.0)
|
11.5 (15.3)
|
Minimum recommended backing pump
for maximum throughput, cfm (m3/hr)
|
17.0 (28.9)
|
30 (51)
|
30 (51)
|
80 (136)
|
Backstreaming rate at inlet flange
mg/cm2/min (standard cold cap)*
|
5 x 10-4
|
5 x 10-4
|
1 x 10-3
|
1.5 x 10-3
|
Warmup time, minutes
|
15
|
15
|
30
|
45
|
Cooldown time, minutes
With quick cool coil, where applicable
|
10
|
25
|
25
|
48
|
Fluid charge
|
1,000 cc
|
1,000 cc
|
3 U.S. qts.
|
5 U.S. qts.
|
Electrical requirements
|
3 ph
50/60 Hz
120/208/240 V
|
3 ph
50/60 Hz
208/240/380/480 V
|
3 ph
50/60 Hz
208/240/380/480 V
|
3 ph
50/60 Hz
240/415/480 V
|
Power, Watts
|
2200
|
4400
|
4400
|
8100/9600
|
Cooling water, U.S. gpm (l/hr)
at 60-80° F (15-26 °C)
|
0.25 (50)
|
0.40 (80)
|
0.40 (80)
|
1.5 (300)
|
|
|