Model Number
|
Specifications
|
Power/Output
|
Features
|
PSG 500-S/502-S
PSG 510-S/512-S
Pirani Standard Gauge
|
*Range: 5X10-4~1000
*Temp.:
80/250℃(Connection)
5~60℃(Operating)
-20~65℃(Storage)
*Vacuum Connection:
DN 16 ISO-KF, 7/16-20 UNF DN16 CF-R, 4 VCR
1/8” NPT, 1/2” Tube.
8VCR, DN 16 ISO-KF
(Long)
DN16 CF-R Long tube
*Filament:
Tungsten & Nickel
|
*Supply: 14~30 VDC.
*Output Signal:
0~10.3V (Voltage range)
1.9~10V(Measurement)
0~0.5V(Error signal)
10 k0hm (Minimum load)
*PSG500 without setpoint and the others with two setpoints.
*Electrical consumption: FCC68/RJ45, 8 poles, male.
*Cable length: ≦100m
*Protection: IP40
|
*Easy push button ATM and HV adjustment.
*Space saving rugged design.
*Ceramic feedthrough for extremly corrosive applications (PSG510 & PSG512)
*Stainless steel measuring cell with metal-sealed feedthrough.
*Logarithmic signal output for easy integration.
*Nickel filament option for corrosive applications.
*Aluminium housing.
|
Model Number
|
Specifications
|
Power/Output
|
Features
|
PVG 500
PVG 500S/502S
Pirani Standard Gauge
|
*Range: 5X10-4~1000
*Temp.:
80℃(Connection)
5~60℃(Operating)
-20~65℃(Storage)
*Vacuum Connection:
KF16
*Filament:
Tungsten & Nickel
*Protection: IP40
*Cable length: ≦100m
|
*Supply: 14~30 VDC.
*Output Signal:
0~10.3V (Voltage range)
1.9~10V(Measurement)
0~0.5V(Error signal)
10 k0hm (Minimum load)
*PVG500 without setpoint and the others with two setpoints.
*Electrical consumption: FCC68/RJ45, 8 poles, male.
|
*Compact, rugged aluminum housing that mounts in any orientation with a logarithmic signal output.
*orientation with a logarithmic signal output.
*Easy push button ATM and High Vacuum (HV) adjustment.
*Nickel filament option.
*Solution for corrosive applications
*Utilize pressure readings to perform critical operations.
|
Model Number
|
Specifications
|
Power/Output
|
Features
|
PSG 050/051
Pirani Standard Gauge
|
*Range:
5x10-4~1000 mbar
*Temp.:
0~40℃(Operating)
80℃(Storage)
110℃(Filament)
|
*Operating units with vacuum gauge controller VGC012 / VGC023
*Filament: Tungsten, Platinium.
*Vacuum connection:
DN 16 ISO-KF, 1/8” NPT.
|
*Tungsten or platinium filament.
*Constant filament temperature.
*PSG050: Aluminium sensing cell with tungsten filament. Excellent temperature compensation.
*PSG050 1/8” NPT: Stainless steel sensing cell with tungsten filament
*PSG051: Stainless steel sensor with
platinium filament and ceramics feedthrough. Well suited for corrosive processes and high levels of water vapour.
|
Model Number
|
Specifications
|
Power/Output
|
Features
|
PSG 100-S/101-S
Pirani Standard Gauge.
|
*Range:
5x10-4~1000 mbar
*Temp.:
80℃(Bakeout)
10~50℃(Operating)
-20~70℃(Storage)
*Filament: Tungsten,Platinum.
|
*Supply: 14.5~36 VDC.
*Output Signal:0~10.6 V
0.66~10V (Range)
1.33.V/Decade(Relation V/Pressure)
*iNTER
*Cable length: 100m
*Flange: Tube AD 1/2”, DN16 ISO-KF, DN 16 CF-R, 8VCR.
|
*Absolute pressure measurement, independent of gas type.
*Excellent long-term and temperature stability.
*Minimal zero drift.
*Corrosion-resistant alumina construction.
|
Model Number
|
Specifications
|
Power/Output
|
Features
|
PCG 400/410
PCG 400-S/410-S
Pirani Capacitance Diaphragm Gauge
|
*Range: 5x10-4~1500 mbar
*Temp.:
80℃(Bakeout)
10~50℃(Operating)
-20~65℃(Storage)
<160℃(Filament)
*Connector: FCC 68, female, 8 pin(shielded).
*PCG400-S with 2 setpoints FCC 68 and PCG410-S with 1 setpoint 9 D-Sub.
|
*Supply: 15~30 VDC. 2.5W
*Output Signal:
0~10.3 V(400/400S)
0~9 V (410/410S)
*Sensor cable length: 100m
*Flange: DN 16 CF-F, 16 ISO-KF,
*Fitting: 8 VCR(Female), 1/8” NPT, 4 VCR(Female), 4 VCR 900 Female.
*Protection: IP40
|
*Gas-type-independent above 100 mbar (75 Torr) - enables safe venting with any gas mixture
*Higher accuracy and reproducibility at atmosphere.
*Independence of mounting position.
*Faster atmosphere detection.
*Compact design.
*One log/linear output signal over the entire range.
|
Model Number
|
Specifications
|
Power/Output
|
Features
|
PCG 550/552/554
Pirani Capacitance Diaphragm Gauge
|
*Range: 5x10-5~1500 mbar
*Temp.:
≦80℃(Bakeout)
10~50℃(Operating)
-20~65℃(Storage)
≦250℃(Long tube)
*Flange: DN 16 ISO KF or Long tube, DN16 CF-F (or 16 CF-R Long tube), DN25 ISO KF,4VCR
*Fitting:8 VCR, 1/8” NPT, 7/16-20 UNF, 29X29 mm,4 VCR 90O.
|
*Supply: 15~30 VDC.
*Output Signal:0~10.23V,
0~8.68V,0~5.659,0~9.05.
*Interface: RS232C(Standard),DeviceNet,Profibus DP, RS485C.
*Electrical connection:
FCC, 8 pin, D-Sub 9 pin/ 15 pin HD male, Hirschmann, 6 pin.
*Cable length:≦100m,
≦30m(RS232C)
*Filament: Tungsten, Nickel, Ceramic coated.
*Protection: IP40.
|
*Gas-type independent above 10 mbar.
*High accuracy and reproducibility at atmosphere.
*Versatile of mounting orientation.
*Available with Tungsten (PCG550) or Nickel (PCG552) filament or with a fully ceramic coated (PCG554) sensor unit for highly corrosive applications.
*Easy to exchange plug & play sensor
element with on-board calibration
Data.
*Eligible output signal for easy integration.
*Optional atmospheric switch, display and digital interfaces.
|
Model Number
|
Specifications
|
Power/Output
|
Features
|
PCG 750 /752
Pirani Capacitance Diaphragm Gauge
|
*Range:
5x10-5~1500 mbar
*Temp.:
80≦℃(Bakeout)
10~50℃(Operating)
-20~65℃(Storage)
*Flange: KF16
|
*Supply: 15~30 VDC. 2 W
*Output Signal:0~10.23 V
0.61~10.23V (Measurement)
1.286 V/decade (Voltage vs. pressure)
>10 k0hm (Load impedance)
*Interface: RS232C.
*Sensor cable length: 500m
*Flange: KF16
*Power connection FCC, 8 pin.
|
*High accuracy and reproducibility at atmosphere.
*Fast atmospheric detection.
*Gas-type independent above 10
mbar.
*Nickel filament option.
* Compact, rugged housing that mounts in any orientation with a logarithmic signal output.
|