Part Number & Picture
|
Descriptions
|
Feature
|
VS Series
|
Automatic Helium Mass-Spectrometer Leak Detectors combine the simplicity of two-button operation with advanced system.
Intelligence to give you easy access to powerful testing capability.
|
1. Easy-to-Use – Two Button Operation.
2. Powerful – Broad Range of Test Methods.
3. Versatile – Multiple System Configurations.
4. Dependable – Robust Design.
5. Truly Portable – Wireless Remote Control/Display.
|
Outline Drawing
|
FRONT (inches) mm
SIDE (inches) mm
|
Part Number
|
Descriptions
|
Pump / pumping speed
|
Volt
|
VS-PR02
|
Portable.
Leak Detector.
|
DS42 Rotary Vane Pump.
2 m3/hr (34 l/m)
|
120V
220V
|
VS-BR15
|
Bench Mount.
Leak Detector.
|
DS302 Rotary Vane Pump.
14 m3/hr (232 l/m)
|
120V
220V
|
VS-MR15
|
Mobile.
Leak Detector.
|
DS302 Rotary Vane Pump.
14 m3/hr (232 l/m)
|
120V
220V
|
VS-BD30
|
Bench Mount.
Leak Detector.
|
TS620 Dry Scroll Pump.
30 m3/hr (500 l/m)
|
120V
220V
|
VS-MD30
|
Mobile.
Leak Detector.
|
TS620 Dry Scroll Pump.
30 m3/hr (500 l/m)
|
120V
220V
|
|
|
|
|
|
|
Technical Specifications
|
Minimum Detectable Leak at
1000 ppm ambient He
|
5 x 10-12 atm cc/sec.
5 x 10-12 mbar l/s.
5 x 10-13 Pa m3/sec helium.
|
Maximum Test Port Pressure
|
13 mbar / 10 Torr / 1330 Pa.
|
Helium Pumping Speed
|
1.8 l/s.
|
Calibration Routine
|
Automated or Manual (Internal or External).
|
Background Suppression
|
Push Button Initiated Auto Zero, and Auto Zero<Zero function.
|
User Interface
|
High Clarity, Color Display, TFT Touch Screen
|
Selectable Languages
|
English, French, German, Japanese, Korean, Mandarin, Spanish.
|
Automated Cycling
|
Programmable Rough Time, Test Time, Reject Set Points
|
Response Time
|
< 0.5 seconds.
|
Set Points
|
Standard , 5 Set points, N/O or N/C, 3 Leak Rate / 1 pressure / 1 Audio.
|
Serial Interface
|
RS-232 optically-isolated, interface @ 9600 baud (DB-9S).
|
Discrete I/O
|
Optically isolated, Outputs: 5-24 volts VDC, Inputs: DB-25S.
|
Wireless Remote Control (optional)
|
2.4 GHz ISM band, 100m (328 ft) range, 4 AA batteries or 5 VDC supply.
|
|